The IQool™ System

Advanced Temperature Management System

Products2019-08-05T16:43:22+00:00

Intelligent. Safe. Simple.

The IQool™ System is the next generation precision surface cooling device that offers high flow rates resulting in optimal heat exchange. The system is user friendly and provides a proactive approach to managing your patients with visible objective data.

BrainCool Screen

Neoprene Silicone Insulating Pads

  • Easily applied and pre-filled

  • Closed fluid system

  • Easy to clean

  • One pad set for the continuum of patient care

  • High flow rate

Customer Support Services

Customer Support Services

Specifications

Therapy Modes Normothermia: cooling and warming
Hypothermia: cooling and warming
Rewarming: cooling and warming
Heater Capacity 2500 BTU/hr 750 Watts
Circulating Fluid (Coolant) Diluted monopropylene glycol
Reservoir Capacity 4.0 liters
Coolant Flow Rate 6 liters per minute
Patient Probe Type YSI 400 Series compatible
Patient Temperature Inputs Patient Temp 1: control, monitor, alarm
Patient Temp 2: monitor, alarm
Patient Temperature Display Range 28°C to 45°C
82.4°F to 113°F
0.1°C / °F increments
Patient Core Temperature Direct Measurement Accuracy output range**:
±0.3°C (35°C to 42°C)
extended output range:
±0.4°C (28°C to 35°C)
±0.4°C (42°C to 45°C)
Including ±0.2°C accuracy of external temperature probe
Patient Temperature Control Range 33°C to 38.5°C
91.4°F to 101.3°F
0.1 °C/°F increments
Coolant Temperature Display Range 1°C to 45°C / 33.8°F to 113.0°F
0.1 °C/°F increments
Coolant Temperature Control Range (Automatic) 4°C to 40°C / 39,2°F to 104°F
High Coolant Temperature Limit 41°C / 105.8°F
1 °C/°F increments
Low Coolant Temperature Limit -1°C / 30.2°F
1 °C/°F increments
Mains Input 100 VAC, 50/60 Hz, 15.0 Amp (nominal)
115 VAC, 50/60 Hz, 15.0 Amp (nominal)
220-230 VAC, 50/60 Hz, 10.0 Amp
Operating Relative Humidity Range 18°C to 27°C / 64.4°F to 80.6°F
15 – 90% RH
Storage/transport Temperature Range -20°C to 50°C / -4°F to 122°F
15 – 90% RH
Dimensions Height: 41 inches (104.1 cm)
Width: 20 inches (50.8 cm)
Depth: 24 inches (61 cm)
Weight Empty: 73 kg / 161 lbs.
Filled: 77 kg / 169 lbs

Specifications

Therapy Modes Normothermia: cooling and warming
Hypothermia: cooling and warming
Rewarming: cooling and warming
Heater Capacity 2500 BTU/hr 750 Watts
Circulating Fluid (Coolant) Diluted monopropylene glycol
Reservoir Capacity 4.0 liters
Coolant Flow Rate 6 liters per minute
Patient Probe Type YSI 400 Series compatible
Patient Temperature Inputs Patient Temp 1: control, monitor, alarm
Patient Temp 2: monitor, alarm
Patient Temperature Display Range 28°C to 45°C
82.4°F to 113°F
0.1°C / °F increments
Patient Core Temperature Direct Measurement Accuracy output range**:
±0.3°C (35°C to 42°C)
extended output range:
±0.4°C (28°C to 35°C)
±0.4°C (42°C to 45°C)
Including ±0.2°C accuracy of external temperature probe
Patient Temperature Control Range 33°C to 38.5°C
91.4°F to 101.3°F
0.1 °C/°F increments
Patient Temperature Control Range 1°C to 45°C / 33.8°F to 113.0°F
0.1 °C/°F increments
Coolant Temperature Control Range (Automatic) 4°C to 40°C / 39,2°F to 104°F
High Coolant Temperature Limit 41°C / 105.8°F
1 °C/°F increments
Low Coolant Temperature Limit -1°C / 30.2°F
1 °C/°F increments
Mains Input 100 VAC, 50/60 Hz, 15.0 Amp (nominal)
115 VAC, 50/60 Hz, 15.0 Amp (nominal)
220-230 VAC, 50/60 Hz, 10.0 Amp
Operating Relative Humidity Range 18°C to 27°C / 64.4°F to 80.6°F
15 – 90% RH
Storage/transport Temperature Range -20°C to 50°C / -4°F to 122°F
15 – 90% RH
Dimensions Height: 41 inches (104.1 cm)
Width: 20 inches (50.8 cm)
Depth: 24 inches (61 cm)
Weight Empty: 73 kg / 161 lbs.
Filled: 77 kg / 169 lbs